Radiography

Ko­dak Di­rect­View DR 9500

The DR 9500 offers the Flexibility of a Dual-Detector System in a Single-Detector System

Arrange appointment

You are welcome to visit us in the time
from 7:00 a.m. to 5:00 p.m.

069 6301-87202

069 6301-5252

Uni­ver­si­täts­kli­ni­kum Frank­furt
In­sti­tut für Dia­gnos­ti­sche und
In­ter­ven­tio­nel­le Ra­dio­lo­gie
Haus 23C Un­ter­ge­schoss
Theo­dor-Stern-Kai 7
60590 Frank­furt am Main

Röntgen-Team

Verantwortlicher OA Dr. med. M. Harth

marc.harth@unimedizin-ffm.de

Teamleitung Elke Marzi

elke.marzi@unimedizin-ffm.de

Kodak9500C-Bogen.jpg

Very modern multifunction system Kodak DR9500 C-arm with flat detector

Technical Specifications:

direct digital x-ray detector
- Detector system made of amorphous Selenium
- Thin-Transistorarray (TFT) for reading the signals
- 35 x 43 cm aktive Detektorfläche
- 2560 x 3072 Detector elements/Pixel
- Edge length of detector element 139µm
- 3,6 Line pairs/mm
- Field depth 14 bit internal
- Display 12 bit log
- 15,8 MB/image
MTF
-> 95% at 1,0 Lp/mm
-> 83% at 2,0 Lp/mm
-> 64% at 3,0 Lp/mm
DQE
-> 44% at 1,0 Lp/mm
-> 37% at 3,0 Lp/mm
-> 26% at 3,0 Lp/mm
- Image per minute=2
-Temperature range: 10 -35C in examination room
- Humidity: 30 - 80% noncondensed