Siemens Magnetom Trio -Tim System 3 Tesla
TIM
- 102 seamlessly integrated Matrix-Coil-Elements and upto 32 HF-channels.
- 50cm FoV. Full body imaging up to 181cm.
- iPAT2. Unlimited PAT up to 16.
Compact Magnet
- Shorter 3T Magnet
- Homogenity: 0.25 ppm VRMS for 40 cm DSV (diameter spherical volume), 1ppm VRMS for 50 cm DSV (diameter spherical volume)
- Spontaneous Quench-Rate = 0
Portrayal of Magnets (Image source: Siemens Medical)
AudioComfort despite strong gradient echo frequencies.
- The integrated hard- and software reduces the noise levels by up to about 20 Decibel dB (A) as compared to standard 3T MRI machines which means a noise reduction up to 90%.
- Stregth of magnetic field (gradient field) up to 45 mT/m (effective 72 mT/m).
- Slew Rate up to 200 T/m/s (effective 346 T/m/s).
- A bigger field of view (FoV) of up to 50 cm for whole body examinations.
- Rapid, compact, water-cooled gradient (RF) amplifier in solid-state-technology for a minimal TR of 1.5 ms and minimal TE of 0.6 ms with a 256 matrix.
Document Date:
unknown
Author:
Jessen Gurung, Marc Harth, Ahmed Koujan
|
Zurück zum Seitenanfang
|
Zur Navigationshilfe