Navigationshilfe

Hauptnavigation

Seiteninhalt

Banner

Digital X-ray Room (Kodak)

The following examinations are performed in this room:

  • All skeletal x-rays
  • Intravenous Pyelogram
  • Chest X-rays
Stativ Kodak 7500

Stativ from Kodak 7500

Technical Specifications:

direct digital x-ray detector
- Detector-System made of amorphous
selenium.
- Thin-Transistorarray (TFT) for reading the signals

- 35 x 43 cm active detector surface

- 2560 x 3072 Detector elements/Pixel

- Edge length of detector element 139µm

- 3,6 Line pairs/mm

- Field depth 14 bit internal
- Display 12 bit log

- 15,8 MB/image

MTF
-> 95% at 1,0 Lp/mm
-> 83% at 2,0 Lp/mm
-> 64% at 3,0 Lp/mm

DQE
-> 44% at 1,0 Lp/mm
-> 37% at 3,0 Lp/mm
-> 26% at 3,0 Lp/mm

- Image per minute=2

-Temperature range: 10 -35C in examination room

- Humidity: 30 - 80% non-condensed

 

Document Date: 2006/10/16   Author: Ahmed Koujan

|  Zurück zum Seitenanfang | Zur Navigationshilfe

Seitenabschlussleiste


Druckversion: 18.05.2012 10:22:59
http://radiologie-uni-frankfurt.de/content/e4864/e1506/e2952/e3482/e627/index_eng.html